Device Problems
Crack (1135); Battery Problem (2885)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
|
Event Description
|
On (b)(6) 2016, the reporter contacted animas, alleging a casing/condition (cracked/damaged casing) issue.The reporter alleged damage to the battery compartment.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the alleged malfunction has the ability to result in a delay in treatment or long term cessation in delivery if the damage impacts the power circuit or cartridge compartment.
|
|
Manufacturer Narrative
|
Follow-up #1: date of submission 08/30/2016.Device evaluation: the device has been returned and evaluated by product analysis on 08/27/2016 with the following findings: investigation revealed cracks in the battery compartment.Unrelated to the original complaint, the audio bolus button cover was damaged; however, the bolus button was responsive during the investigation.Evidence of moisture was observed in the battery compartment; a leak test failed due to a battery compartment leak.The pump was opened and no further evidence of moisture was found internally.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
|