The patient was placed on an extracorporeal circulatory support device in conjunction with extracorporeal membrane oxygenation on (b)(6) 2016.It was reported that a short time after bedside ecmo cannulation, the bedside nurses smelled smoke, at which time the console shut down and the speed went to zero.The patient immediately decompensated with desaturation and hypotension.The nurses emergently clamped the circuit and switched to the backup console and motor.The circuit was switched to a 2nd generation circuit system.The patient immediately stabilized once flow was initiated.It was reported that no alarms were noted at the time of the incident.
|
Device evaluation: the report of burnt smell emitting from the returned primary console was confirmed.Two printed circuit board (pcb) components were found to be defective.Based on the manufacturer¿s complaint experience, the motor was suspected to be responsible for the damaged pcbs; however, during the analysis of the suspect motor, no device related issues were detected.Therefore, the root cause could not be could not be conclusively determined.Once the damaged pcb components were replaced and the returned primary console was subjected to the reprocessing and maintenance procedure, the unit passed all tests.The returned motor remained suspect and was therefore scrapped as a precaution.A review of the device history records revealed no deviations from manufacturing or quality assurance specifications.The manufacturer is closing the file on this event.
|