• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD INC. PORTEX CUSTOM SINGLE LIMB CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD INC. PORTEX CUSTOM SINGLE LIMB CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) Back to Search Results
Catalog Number C1110
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Customer has not yet returned the device to the manufacturer for device evaluation.When and if the device becomes available and is returned and evaluated the manufacturer will file a follow-up report detailing the results of the evaluation.
 
Event Description
The customer reported disconnecting or loose connection points resulting in a leak alarm on the portex single limb circuit.The issues were noticed nearly immediately.No patient or clinical injury reported.Additional information has been requested.
 
Manufacturer Narrative
One portex® custom single limb circuit was returned for investigation.The device was received inside a plastic bag with its original closed packaging.The customer reported that two devices contributed to two reported events (one device per event).The customer returned one device for evaluation.It could not be determined which reported occurrence that device was associated with; therefore, the evaluation of the device will be used for the medwatch.The following mfr were submitted related to the evaluation: 2183502-2016-01667 and 2183502-2016-01668.Visual inspection found no deformations/holes along the tube and the device was correctly assembled; no defects were observed.Additionally, no holes were found on the breathing bag.During functional testing, the returned tube and bag were connected to the leak tester device; no leaks were detected.The returned device then underwent a retention and bond strength test using a calibrated weight and brackets.During the strength and bond test, no separation on the parts or on the bonding was observed.After the strength and bond test, the tube and bag were re-connected to the leak testing device; no leaks were observed.Investigation was unable to find fault with the returned device and found that the device operated as intended.Mfr# clarification: new registration number (b)(4) ((b)(4)) has been received, however, this initial mdr was filed under the prior registration number (b)(4) ((b)(4)).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX CUSTOM SINGLE LIMB CIRCUITS
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL ASD INC.
1265 grey fox road
st. paul MN 55112
Manufacturer (Section G)
SMITHS HEALTHCARE MFG
s.a. de c.v. ave calidad no. 4
parque industrial internaciona
tijuana 55425
MX   55425
Manufacturer Contact
lisa perz
1265 grey fox rd
st. paul, MN 55112
7633833074
MDR Report Key5848451
MDR Text Key51224872
Report Number2183502-2016-01667
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 07/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC1110
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/12/2016
Initial Date FDA Received08/04/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/02/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-