• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number 250
Device Problems Computer Software Problem (1112); Looping (1370); Failure to Align (2522)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
It was reported the physician's handheld programming computer was stuck in the screen alignment loop.The issue could not be resolved by a hard reset.Product return is expected, but the device has not been received by the manufacturer to date.
 
Event Description
Product analysis (pa) for the returned handheld programmer (hhd) showed the reported allegation of a mechanical problem/hardware failure was not verified.Note the original ac adapter and serial cable were not returned.No anomalies associated with the hhd performance were noted during testing using a known good ac adapter or the main battery with a full charge.The hhd performed according to functional specifications.Pa for the software showed there were no anomalies associated with the flashcard software or databases.The flashcard and software performed according to functional specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5851060
MDR Text Key52014531
Report Number1644487-2016-01759
Device Sequence Number1
Product Code MUZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number250
Device Lot Number1063824
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2016
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 07/07/2016
Initial Date FDA Received08/05/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received09/01/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/11/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-