Model Number 8070
Device Problem Break (1069)
Patient Problem Fall (1848)
Event Type  malfunction  
Manufacturer Narrative
Evaluation of the returned bed confirmed that one tooth from the left side patient access panel was missing.The malfunction was repaired and returned back to the customer for use.The customer further reported that the patient that fell required stitches for a cut they received on the lip.Posey beds are multi-use, serviceable items.As such, it is anticipated that the units may occasionally require repair, and as part of standard care the beds should be inspected prior to use.If damage is noted during these routine bed inspections, the unit should not be put into use with a patient and should be returned to posey for servicing.If a zipper element is missing, it can allow for the unintentional opening of the canopy window.Applying pressure directly to the unsecured area will reveal the breach, which is why the user manual advises caregivers to apply direct pressure along the entire length of the zipper.Although the cause of the missing element cannot be confirmed, it is possible that excessive use or normal wear and tear could have attributed to the damage.Of note, the canopy was over 24 months old when it was received in for service.Service issues are trended and reviewed by management on a monthly basis.As part of this monthly review, trends and excursions above control limits will be assessed, documented and acted upon as warranted.No corrective or preventative actions are necessary at this time.(b)(4).
 
Event Description
Customer reported the zipper teeth are not-aligned and the zipper is jagged throughout the zipper tape.The spacing between teeth are torn and pulled out of place.The panel zipper opened up and the patient fell out of the canopy.The customer confirmed the patient suffered injuries to his lip requiring stitches to his lip.The exact date is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POSEY BED 8070
Type of Device
PATIENT BED WITH CANOPY/RESTRAINTS
Manufacturer (Section D)
J.T. POSEY COMPANY
5635 peck rd
arcadia CA 91006
Manufacturer (Section G)
POSEY, S. DE R.L. DE C.V.
ave. ferrocarril no. 16901. bo
colonia rio tijuana, 3ra. etap
tijuana 22664
MX   22664
Manufacturer Contact
william hincy
posey company
5635 peck road
arcadia, CA 91006
6264433143
MDR Report Key5851152
MDR Text Key51296790
Report Number2020362-2016-00025
Device Sequence Number1
Product Code OYS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK103817
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Service Personnel
Type of Report Initial
Report Date 07/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other Caregivers
Device Model Number8070
Device Catalogue Number8070
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/21/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/11/2016
Initial Date FDA Received08/05/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/22/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1