(b)(4).(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
It was reported from (b)(6) that the attachment device had an undetermined malfunction.During the pre-repair diagnostics assessment, it was observed that the bearings had worn components.It was further determined that the bearings and sleeve/case were overheating; resulting in slight vibrations during the e-test.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|