• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO ADAPTA DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO ADAPTA DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number ADDR03
Device Problems Device Expiration Issue (1216); Pacing Problem (1439)
Patient Problem Bradycardia (1751)
Event Date 06/10/2016
Event Type  Injury  
Manufacturer Narrative
This device was reported as included in the field action.Based on the information received and without the return of the product, it cannot be confirmed that this device performed as described in the field action.It is included in the field action in the abundance of caution.
 
Event Description
It was reported that the patient presented to the emergency room with symptoms of weakness and felling faint due to bradycardia.A device check indicated that the device had switched to ventricular pacing only due to elective replacement indicator (eri).The patient only had an atrial lead as the ventricular port was plugged at implant; so consequently, the patient now was not being paced at all.The device was explanted and replaced.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.This device was reported as included in the field action noted but returned product investigation found the device did not perform as described in the field action.The device is no longer included as part of the field action.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADAPTA DR
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key5859562
MDR Text Key51692608
Report Number3004209178-2016-15968
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 06/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/14/2009
Device Model NumberADDR03
Device Catalogue NumberADDR03
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/15/2016
Initial Date FDA Received08/09/2016
Supplement Dates Manufacturer ReceivedNot provided
08/23/2016
Supplement Dates FDA Received10/06/2016
09/23/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/24/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-2120-2011
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age36 YR
-
-