• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - FREMONT (CE) ANGIOJET® ULTRA SYSTEM CONSOLE; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - FREMONT (CE) ANGIOJET® ULTRA SYSTEM CONSOLE; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number 105650-037
Device Problems Device Remains Activated (1525); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/26/2016
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: over 18 years device evaluated by mfr: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
 
Event Description
It was reported that the system could not be stopped when aspirating by the foot pedal.An angiojet ultra system console was selected for a venous thrombectomy procedure.The first part of the procedure was normal.After aspirating for 80 seconds, the foot pedal was released to prepare to stop aspiration.However, at this time, the system could not be stopped and it was still in running status.The keyboard was not responding, so the physician tried to shut down the system with the power button.The system was restarted and it changed the language from chinese to english and showed that the foot pedal was on.The console display showed: change the language.At this time only the power and alarm reset could be used.Then the physician changed the language and clicked "ap" after which the console showed "the aspiration volume 200cc, start the aspiration when the foot petal was off." the system could not be used and was shut down again.After one hour, the physician restarted the system.The system automatically switched to the priming procedure and showed an error message to "connect the saline supply." the pump cabinet could not be opened and the catheter could not be removed.After two hours, the physician restarted the system and the system automatically ran.The pump cabinet was opened and the catheter was removed.After the catheter was removed, the system displayed a message that said "illegally installed the catheter." the physician shut down the system and no further actions were taken.The procedure was not completed due to this event.There were no patient complications and the patient condition was noted as stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOJET® ULTRA SYSTEM CONSOLE
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC - FREMONT (CE)
47215 lakeview blvd
north dock
fremont CA 94538
Manufacturer (Section G)
BOSTON SCIENTIFIC - FREMONT (CE)
47215 lakeview blvd
north dock
fremont CA 94538
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5866782
MDR Text Key51827161
Report Number2134265-2016-07461
Device Sequence Number1
Product Code MCX
Combination Product (y/n)N
Reporter Country CodeCN
PMA/PMN Number
P980037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 08/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number105650-037
Device Catalogue Number105650-037
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/01/2016
Initial Date FDA Received08/10/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/17/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-