(b)(4).(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to excessive wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
It was reported by (b)(6) that during service and evaluation, it was discovered that the locking mechanism was worn out and not functioning on the motor device.It was further determined that the motor was worn out.It was further determined during the pre-repair diagnostic assessment that the device failed for noise and safety assessments.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|