• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS 5.0CM SHORT ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS 5.0CM SHORT ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT Back to Search Results
Catalog Number SHORT
Device Problems Break (1069); Detachment Of Device Component (1104); Component Missing (2306); Protective Measures Problem (3015)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Udi: gtin is unavailable as the product was made prior to compliance date; (b)(4).(b)(6).The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.However, the assignable root cause was not determined.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported from (b)(6) that during service and evaluation, it was observed that the short attachment device had damaged component: bearings.It was noted that the bearings have fallen, the balls and cage were missing, and the extension sleeve was damaged.It was also noted that the device failed pre-repair assessment tests for cutter insertion.It was noted in the service order "bearing damaged".This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.0CM SHORT ATTACHMENT
Type of Device
MOTOR, DRILL, ELECTRIC - ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key5879658
MDR Text Key53063781
Report Number1045834-2016-12459
Device Sequence Number1
Product Code HBC
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K011444
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 05/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSHORT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/07/2016
Initial Date FDA Received08/16/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/26/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-