IRIS INTERNATIONAL IQ 200 SYSTEM; AUTOMATED URINALYSIS SYSTEM Back to Search Results
Catalog Number 700-3347
Device Problem Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/02/2016
Event Type  malfunction  
Manufacturer Narrative
The field service engineer (fse) was at the customer site but the customer had resolved the issue before his arrival.The cause for the reported leaking is attributed to a ruptured tubing that connects the sample filter and the pipette bypass valve (pbv).The customer replaced the tubing and that resolved the reported issue.The customer ran controls and the passed without any issue.Bec internal identifier for this report is (b)(6).
 
Event Description
The customer reported a leak on their iq200 system.The leak was observed on the left side of the instrument and was not contained.The customer was wearing personal protective equipment (ppe) consisting of a lab coat and gloves.There was no exposure to the leaking hazardous fluid.Erroneous patient results were not generated and there was no change or effect to patient treatment in connection to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IQ 200 SYSTEM
Type of Device
AUTOMATED URINALYSIS SYSTEM
Manufacturer (Section D)
IRIS INTERNATIONAL
9172 eton ave
chatsworth CA 91311
Manufacturer (Section G)
IRIS INTERNATIONAL
9172 eton ave
chatsworth CA 91311
Manufacturer Contact
gopal mohanty
9172 eton avenue
chatsworth, ca, CA 91311
8185277379
MDR Report Key5883993
MDR Text Key52390425
Report Number2023446-2016-00338
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K022774
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number700-3347
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/02/2016
Initial Date FDA Received08/17/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/17/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown