It was reported that ¿several¿ patients have experienced significant bradycardia due to no flow or restricted flow occurring during infusions in which a clearlink tubing set was used.The tubing sets were being used with unspecified pumps.It was reported that when an infusion was being run at a low rate, the pump alarm ¿does not go off until late¿ and therefore, does not sense early on if the infusion is running correctly resulting to patient not getting medication.As a result this has resulted in patient reactions such as significant bradycardia and clinicians not being able to inject drugs into the patients (no further detail was provided).No further information could be obtained such as how many patients this occurred with, what medical interventions were required, what drug was being infused and/or the patients outcomes from the events.No additional information is available.
|