Model Number 250
Device Problems Battery Problem (2885); Charging Problem (2892)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/20/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported by the company representative, while at the physician's office, that the physician's tablet would not charge.An alternate power cable was used and the tablet was allowed to charge for 20 minutes.She reported that she saw red lights appear when the tablet was charged with an alternate charger and made the assumption that the tablet charger was faulty.It was later reported that the company representative was able to use the tablet while it was plugged into the wall, but it would not turn on at all after being unplugged.The company representative stated she would return the tablet but a replacement was not needed.It was also clarified there were no issues with the power cables and there were no reports of the tablet being dropped or anything of that nature which could have caused an issue.
 
Event Description
Product analysis (pa) for the returned tablet found the reported allegation of ¿battery failure, will not hold charge¿ was verified.During analysis, it was identified that the main battery was defective and no longer able to hold a charge and power the tablet.Once the battery was replaced with a known good battery, no further anomalies were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5885207
MDR Text Key52503662
Report Number1644487-2016-01864
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number250
Other Device ID NumberVERSION 11.0.5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2016
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 07/20/2016
Initial Date FDA Received08/18/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received09/01/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/27/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-