MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065751763
Device Problems Unstable (1667); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/25/2016
Event Type  Injury  
Manufacturer Narrative
A sample has been received and in-house testing is in progress.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported that the patient experienced an unstable chamber during a procedure.A vitrectomy was performed.Additional information has been requested, but none has been received to date.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
The clinical analyst reviewed this file and stated the following: ¿a customer reported that the patient experienced an unstable chamber during a procedure.A vitrectomy was performed.Additional information was requested, but none has been received to date.The company service representative examined the system and no problems were found.The system was then tested and met all product specifications.Posterior capsule (pc) tear is a potential consequence of cataract extraction by phacoemulsification.Predisposition to pc tear or zonular dehiscence can be influenced by many factors including, but not limited to, congenital posterior lenticonus, posterior sub capsular (psc) cataract, poor visibility secondary to patients comorbidity [ie.Dense arcus, pterygium, band keratopathy, corneal scars, interstitial keratitis], poor microscope illumination [red reflex], ergonomic obstacles, limited intraocular working space, abnormally long or short axial lengths, pseudoexfoliation, zonular laxity, poor dilation, intraoperative floppy iris syndrome (ifis), dense cataracts, asteroid hyalosis, or inadvertent patient movement.The conditions that increase the risk of pc tear during phacoemulsification include ergonomic obstacles, limited intraocular working space, poor visualization, increased nuclear size and density, weakened zonule, a radial tear in the capsulorhexis, and an inability to rotate the nucleus or epinucleus.These conditions may arise either because of the ocular anatomy, or because of surgical technique.There is no evidence contained within the reported information at this time that indicates that the design or performance of the system contributed to the event.¿ the system was examined and the reported event was not replicated.The system was removed from the facility and returned to alcon for further testing.A review of the customer¿s complaint history for the last 24 months did not show any previous complaints of this kind against the system.Based on qa assessment, the product met specifications at the time of release.The system was received for evaluation.A visual assessment of the returned sample revealed no non-conformities.The system was powered on without nonconformity.The fluidics portion was then performed and the system was found to meet specifications.Posterior capsule tear is an issue that is occasionally reported with cataract surgery.However, a review of the complaint trends shows that the frequency reported is within known levels for this event.A root cause cannot be determined conclusively.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
rita lopez
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514846
MDR Report Key5888855
MDR Text Key52525396
Report Number2028159-2016-03699
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 11/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number8065751763
Other Device ID NumberASKU
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/17/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/25/2016
Initial Date FDA Received08/19/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received09/09/2016
11/21/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/06/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-