The clinical analyst reviewed this file and stated the following: ¿a customer reported that the patient experienced an unstable chamber during a procedure.A vitrectomy was performed.Additional information was requested, but none has been received to date.The company service representative examined the system and no problems were found.The system was then tested and met all product specifications.Posterior capsule (pc) tear is a potential consequence of cataract extraction by phacoemulsification.Predisposition to pc tear or zonular dehiscence can be influenced by many factors including, but not limited to, congenital posterior lenticonus, posterior sub capsular (psc) cataract, poor visibility secondary to patients comorbidity [ie.Dense arcus, pterygium, band keratopathy, corneal scars, interstitial keratitis], poor microscope illumination [red reflex], ergonomic obstacles, limited intraocular working space, abnormally long or short axial lengths, pseudoexfoliation, zonular laxity, poor dilation, intraoperative floppy iris syndrome (ifis), dense cataracts, asteroid hyalosis, or inadvertent patient movement.The conditions that increase the risk of pc tear during phacoemulsification include ergonomic obstacles, limited intraocular working space, poor visualization, increased nuclear size and density, weakened zonule, a radial tear in the capsulorhexis, and an inability to rotate the nucleus or epinucleus.These conditions may arise either because of the ocular anatomy, or because of surgical technique.There is no evidence contained within the reported information at this time that indicates that the design or performance of the system contributed to the event.¿ the system was examined and the reported event was not replicated.The system was removed from the facility and returned to alcon for further testing.A review of the customer¿s complaint history for the last 24 months did not show any previous complaints of this kind against the system.Based on qa assessment, the product met specifications at the time of release.The system was received for evaluation.A visual assessment of the returned sample revealed no non-conformities.The system was powered on without nonconformity.The fluidics portion was then performed and the system was found to meet specifications.Posterior capsule tear is an issue that is occasionally reported with cataract surgery.However, a review of the complaint trends shows that the frequency reported is within known levels for this event.A root cause cannot be determined conclusively.(b)(4).
|