SMITH & NEPHEW, INC. TRUCLEAR INCISOR DEVICE; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Catalog Number 72202536
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/26/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during a hysteroscopy procedure, the window lock on the device came out twice during the procedure.The procedure was completed with this device.No patient injury or complications were reported.
 
Manufacturer Narrative
A visual inspection was performed and showed the blade has been used in the field.The investigation results indicated the window lock laser mark was in the correct location.There is evidence that the blade is bent.This is caused by excessive force applied to the outer tube and would cause an issue with the window lock.No manufacturing related defects were observed.After the evaluation, the root cause for the reported issue was determined to be user error.A review of the device history record was performed which confirmed no inconsistencies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUCLEAR INCISOR DEVICE
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer (Section G)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer Contact
james gonzales
150 minuteman road
andover, MA 01810
5123585706
MDR Report Key5889577
MDR Text Key52616757
Report Number3003604053-2016-00042
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103389
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2019
Device Catalogue Number72202536
Device Lot NumberE91985
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/26/2016
Initial Date FDA Received08/19/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received09/26/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/21/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-