• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KAVO DENTAL GMBH GENTLEPOWER LUX CONTRA ANGLE 25LPA; DENTAL CONTRA ANGLE HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KAVO DENTAL GMBH GENTLEPOWER LUX CONTRA ANGLE 25LPA; DENTAL CONTRA ANGLE HANDPIECE Back to Search Results
Model Number 25 LPA
Device Problems Overheating of Device (1437); Dent in Material (2526)
Patient Problem Burn, Thermal (2530)
Event Date 08/01/2016
Event Type  malfunction  
Manufacturer Narrative
During a short test run the head heated up quickly.During analysis it was found that the head was bent in around the push button.This causes the push button to stick in causing unusual inner friction and therefore heat up of the head.Root cause is that the head received external hits, e.G.Dropping during the reprocessing.To avoid such issues the user instruction contains already several notes, warnings and requests how to prepare the handpiece for each treatment and how to use it: warning: hazards for the care provider and the patient.In the case of damage, irregular running noise, excessive vibration, untypical warming or when the cutter or grinder cannot be held.Do not use further and notify service.Caution: burning hazard from hot instrument head or hot instruments cover.If the instrument overheats, burns may arise in the oral area.Never contact soft tissue with the instrument head or instrument cover the following guidelines must be observed to ensure save use of the electrically driven contra-angle handpieces: the service instructions for contra-angle handpieces must be precisely following when using kavo spray or quattrocare care systems.Before each use, the contra-angle handpiece must be checked for external damage.Before each use, perform a test run with the contra-angle handpiece, and watch for atypical heating and unusual noise and vibration.Immediately stop using contra-angle handpieces that act unusual.Never press the push button during operation.This also includes lifting the cheek or tongue! to ensure proper function, the medical device must be set up according to the reprocessing methods described in the kavo instructions for use, and the care products and care systems described therein must be used.Kavo recommends specifying a service interval at the dental office for a licensed shop to clean, service and check the functioning of the medical device.This service interval depends on the frequency of use and should be adjusted accordingly.
 
Event Description
While working on tooth# l, patient received a burn mark that spread from inside of lower lip to come up onto lower lip.Burn area 5mm x 10mm in size.No ointment or medication was given to patient for burn.Mother told dental assistance she gave her son over the counter tylenol.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GENTLEPOWER LUX CONTRA ANGLE 25LPA
Type of Device
DENTAL CONTRA ANGLE HANDPIECE
Manufacturer (Section D)
KAVO DENTAL GMBH
bismarckring 39
biberach / riss, bw 88400
GM  88400
Manufacturer (Section G)
KAVO DENTAL GMBH
bismarckring 39
biberach / riss, bw 88400
GM   88400
Manufacturer Contact
klaus reisenauer
bismarckring 39
biberach / riss, bw 88400
GM   88400
735156
MDR Report Key5892786
MDR Text Key52618737
Report Number3003637274-2016-00056
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K073478
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Dentist
Type of Report Initial
Report Date 08/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Dentist
Device Model Number25 LPA
Device Catalogue Number1.002.1526
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/10/2016
Initial Date FDA Received08/22/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/03/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age5 YR
-
-