• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY MICROSTAAR® INJECTOR; INTRAOCULAR LENS GUIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY MICROSTAAR® INJECTOR; INTRAOCULAR LENS GUIDE Back to Search Results
Device Problem Entrapment of Device (1212)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/13/2016
Event Type  malfunction  
Manufacturer Narrative
Pt information: unk.Event date: unk.Expiration date - unk.Implant and/or explant dates: na.Initial reporter: unk.Device evaluated by manufacturer? no.Product not returned.Mfg date: unk.(b)(4).Product not returned.
 
Event Description
The reporter indicated an aa4203tl silicone single piece lens, 21.0 se/2.0 diopter, became jammed in the injector.There was no patient contact.The reporter indicated there was an injector problem.Additional information has been requested but none has been forthcoming.If additional information is received, a supplemental medwatch report will be submitted.
 
Manufacturer Narrative
The product for this complaint was not returned for evaluation, however, the lens was returned.Visual inspection of the lens found pieces of the lens optic and one haptic torn off and missing.The lens was returned dry and there was evidence of clear surgical residue.The cartridge was not returned.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROSTAAR® INJECTOR
Type of Device
INTRAOCULAR LENS GUIDE
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
michelle andres
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key5898596
MDR Text Key52840455
Report Number2023826-2016-01156
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K954600
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/25/2016
Initial Date FDA Received08/23/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/09/2016
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-