• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS EMAX2PLUS; MOTOR, DRILL, ELECTRIC - HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS EMAX2PLUS; MOTOR, DRILL, ELECTRIC - HANDPIECE Back to Search Results
Catalog Number EMAX2PLUS
Device Problems Defective Device (2588); Device Operates Differently Than Expected (2913); Protective Measures Problem (3015)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause could not be determined.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported by (b)(6) that during service and evaluation, it was observed that the motor device had damaged component - bearings and defective drivers from motor and drive shaft.It was further determined that the device failed safety assessment at pretest.It was noted in the service order that the motor in the handpiece did not run as it should.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMAX2PLUS
Type of Device
MOTOR, DRILL, ELECTRIC - HANDPIECE
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key5899636
MDR Text Key53548706
Report Number1045834-2016-12515
Device Sequence Number1
Product Code HBC
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K080802
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 07/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEMAX2PLUS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/29/2016
Initial Date FDA Received08/24/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/10/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-