The product analysis indicates that part number 8227410 (paired electrodes) from lot number 0211358568 were received.The paired electrodes were opened and removed from their pouch.The ground and stimulation return electrodes were still in the sealed pouch and therefore not tested.A microscope and multimeter were used for testing.When compared to the assembly drawing, visually there was no external damage to the electrodes which would have resulted in the reported event, which likely rules out mishandling.When compared to the assembly drawing for paired subdermal electrodes, end to end resistance for each side of the paired electrodes shall be <2.0 ohms and infinite ohms between needle tips [no short circuit].The actual measurements of the red paired were as follows - both poles measured 1.5 ohms with no short circuit between poles (in specification).The actual measurements of the blue paired were as follows - both poles measured 1.6 ohms however measuring from needle to needle showed 0.4 ohms which indicates a short circuit between poles (out of specification) and is likely to have resulted in the intermittent behavior that the customer experienced.
|