The endowrist vessel sealer instrument was returned for evaluation.Failure analysis did not confirm the customer reported complaint of the vessel sealer instrument sealing intermittently.The instrument delivered energy with no issues at all angles.The electrical continuity test and gap test passed.Grip test was also performed and passed.The instrument was placed on an in house da vinci system and passed the engagement and recognition test.The cut test passed and energy was delivered.Minor debris was found at the instrument tip and there was no snake wrist or conductor wire damage.Based on the information provided, this complaint is being reported due to the following conclusion: during a da vinci assisted colon resection procedure, the vessel sealer instrument did not seal the patient's tissue.While there was no report of any patient harm, recurrence of the reported failure mode could cause or contribute to an adverse event if the failure mode was to recur.It is unknown what caused the sealing issue.
|
It was reported that during a da vinci assisted colon resection procedure, the endowrist vessel sealer was working and then stopped sealing.There was no report that any fragment(s) from the instrument fell into the patient and there was no report of any patient harm, adverse outcome or injury due to the sealing issue.The planned surgical procedure was completed.
|