MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY FLEXTOME¿ CUTTING BALLOON¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY FLEXTOME¿ CUTTING BALLOON¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H749RB4400100
Device Problems Material Rupture (1546); Device Damaged Prior to Use (2284)
Patient Problem No Patient Involvement (2645)
Event Date 08/08/2016
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.(b)(4).Device evaluated by mfr.: the device was not returned for evaluation.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
 
Event Description
It was noted that balloon rupture occurred.A 10/4.00 flextome¿ cutting balloon¿ was selected for use.Upon preparation, it was noted that the balloon ruptured.The procedure was completed with a different device.There were no patient complications and the patient's condition was good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXTOME¿ CUTTING BALLOON¿
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5917691
MDR Text Key53592119
Report Number2134265-2016-07739
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
PMA/PMN Number
P950020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 08/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/29/2018
Device Model NumberH749RB4400100
Device Catalogue NumberRB440010
Device Lot Number0018680466
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/08/2016
Initial Date FDA Received08/31/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/21/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-