MAUDE Adverse Event Report: STRYKER GMBH INTRAMEDULLARY ARTHRODESIS IMPLANT SMART TOE II 16MM / 10° ANGLE FOR PIP ARTHRO; PIN, FIXATION, SMOOTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH INTRAMEDULLARY ARTHRODESIS IMPLANT SMART TOE II 16MM / 10° ANGLE FOR PIP ARTHRO; PIN, FIXATION, SMOOTH Back to Search Results
Catalog Number ST0A16P
Device Problem Break (1069)
Patient Problem Complaint, Ill-Defined (2331)
Event Date 08/11/2016
Event Type  Injury  
Manufacturer Narrative
Device remains implanted.Additional information was requested and if received will be provided on a supplemental report.Device remains implanted.
 
Event Description
It was reported by the surgeon that the patient had 2 stryker smart toes break.The breakage was confirmed on x-ray.The broken devices were not explanted.
 
Manufacturer Narrative
The reported event that intramedullary arthrodesis implant smart toe ii 16mm / 10° angle for pip arthro was alleged of 'implant breakage after surgery' could be confirmed.Based on investigation, the root cause was attributed to be patient related.The failure was caused by a non-union.The patient x-ray shows 2 broken smarttoe implants in the left foot of the patient: 1 broken smarttoe in the 2nd ray and 1 broken smarttoe in the 3rd ray.Both arthrodesis did not occur and there is no bony consolidation.An internal fixation with smarttoe as well as any kind of internal fixation of small bones (e.G.With k-wires) bears a significant number of risks in the healing period.To a lesser extend these risks are device related.The majority of the risks are caused by poor bone quality, poor blood supply of the fragments poor surgical technique and/or poor patient compliance.Typical risks are: delayed union, non-union, malunion, and pain.Some of these risks may be mild and would not require specific therapeutic measures (s2).Some of them might require specific measures (e.G.Additional bracing in delayed union or non-union), specific medication (e.G.Antiphlogistics and/or analgetics) (s3) or would require unintended revision surgery (e.G.Corrective procedures in the case of malalignment or malunion or revision of a non-union) (s3).Please note that the operative technique (st2-st-1_smart_toe_op_tech_2015-7181) reads: ''precautions: smarttoe implants are not intended for immediate postoperative weight bearing.Be sure that the postoperative loading of the internal fixation is reduced to a minimum (e.G.With application of a forefoot off-loading shoe) until bone consolidation is confirmed by follow up x-ray examination (normally after 4 - 6 weeks).[.] postoperative care: smart toe implants are not intended for immediate postoperative weight bearing.Be sure that the postoperative loading of the internal fixation is reduced to a minimum (e.G.With application of a forefoot off-loading shoe) until bone consolidation is confirmed by follow up x-ray examination (normally after 4-6 weeks).[page 14]'' [original statement(s)] moreover, the instruction for use (v15066 rev b smarttoe xfuse ifu ((b)(4))) reads: ''the product does not allow the immediate resumption of activity by the patient and is not designed to support an immediate load.Immobilization is necessary during the osteosynthesis.[¿] - it is necessary to inform the patient of the precautions to be taken to ensure the success of the implantation.'' [original statement(s)] a review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event Description
It was reported by the surgeon that the patient had 2 stryker smart toes break.The breakage was confirmed on x-ray.The broken devices were not explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRAMEDULLARY ARTHRODESIS IMPLANT SMART TOE II 16MM / 10° ANGLE FOR PIP ARTHRO
Type of Device
PIN, FIXATION, SMOOTH
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH   2545
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5922590
MDR Text Key53748587
Report Number0008031020-2016-00431
Device Sequence Number1
Product Code HTY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112197
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/31/2020
Device Catalogue NumberST0A16P
Device Lot NumberJ20890
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/11/2016
Initial Date FDA Received09/01/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/13/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/12/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-