A review of the device history records of the specimen device lot does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.The safari guidewire was received for analysis.As received, the specimen consisted of one each clinically used, damaged safari wire 300cm sml crv; returned coiled, loose and double-bagged within "zip-lock" style poly biohazard pouches.The specimen presented numerous offset/mis-aligned/overlapping coil wraps resulting in localized oversized diameters with scraped ptfe coating and coating removal in the overlapping coil region at approximately 14.9cm from the proximal end and at the outside surface of the bend damage.The specimen also presented several bends of varying severity.The coating removal appears consistent with the coil wire rubbing on the inner lumen wall surface of the catheter.Deposits of dried blood-like material are present on and between the coil wraps.There is no indication of the device becoming "decoiled." at this time, it is not possible to assign a definitive root cause for the event as reported.Based on the evidence presented by the sample and the information provided by the supporting documentation, clinical and/or procedural factors appear to have contributed to the event as reported.
|
After system withdrawal the guidewire proximal portion got stuck inside the system catheter, physician experienced high resistance to retrieve the wire.After several attempts the wire was finally removed, further inspection showed a small portion of the wire getting decoiled.Valve implantation was completed normally with this device, patient condition is stable.
|