TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 801558 |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/17/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The only component that was damaged was the lamp shade, melting due to excessive heat produced by the incorrect halogen bulb.The bulb has been discarded.The damaged lamp shade is on the biomed's desk.A replacement lampshade has been ordered.As soon as a replacement is received, the biomed will return the damaged one.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the lamp head on the perfusion system melted, as a wrong light bulb was installed.The device was not changed out, as they are no longer using the lamp, using the base as normal.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
Per follow up with the user facility biomedical technician (bmet), no device will be returned to the manufacturer for evaluation.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Manufacturer Narrative
|
If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|