Catalog Number 05.001.010
Device Problems Device Inoperable (1663); Defective Device (2588); Device Operates Differently Than Expected (2913); Moisture or Humidity Problem (2986)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to improper maintenance, which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported by (b)(6) that during service and evaluation, it was observed that the control unit on the electric pen drive device was not functioning and defective.It was further determined that the device ran clockwise in reverse position.It was further determined that the electronic control unit (ecu) was faulty and had water stains on motor surface.It was also noted that the device failed pre-test for direction of rotation, function with test hand switch and function of foot pedal.It was noted in the service order that the device was not functioning when the cable was attached to the console.It was noted that it was tested with different cable and console device but the issue was not resolved.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPD 60000 RPM
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
mark vornheder
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5937746
MDR Text Key54752918
Report Number8030965-2016-14793
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K043310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 08/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/26/2016
Initial Date FDA Received09/08/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/17/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N