• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON CONCHA NEPTUNE; RESPIRATORY GAS HUMIDIFIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON CONCHA NEPTUNE; RESPIRATORY GAS HUMIDIFIER Back to Search Results
Catalog Number 425-00
Device Problems No Display/Image (1183); Power Problem (3010)
Patient Problem No Patient Involvement (2645)
Event Date 08/30/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A visual, functional, and dimensional inspection of the product involved in the complaint could not be conducted since the product was not returned.A device history record review was conducted no issues were found related to the complaint.A document assessment (fmea) was conducted and no changes required.Corrective actions cannot be established since it is necessary to receive the physical sample to perform a proper investigation and confirm the alleged defect.At this time since the device sample is not available, it is not possible to determine the source of the defect reported.Customer complaint cannot be confirmed since the device sample is not available to perform a proper investigation and determine the root cause.If the device sample becomes available at a later date, this complaint will be updated accordingly.
 
Event Description
The customer alleges that the device will not power on and the high and low priority alarm indicators (yellow and red lights) will not turn on or off or flickers.
 
Manufacturer Narrative
(b)(4).The sample was returned for evaluation.A visual exam was performed and there were no defects observed.Functional testing was performed and the unit was connected to 110vac.The unit passed the initial power connect test and navigated through the power-on self-test (p.O.S.T.) with no issues.During the temperature display accuracy test the unit displayed the correct temperatures and properly alarmed in the high temperature scenario.Based on the investigation performed, the reported complaint could not be confirmed.Functional testing did not reveal any operational anomalies.
 
Event Description
The customer alleges that the device will not power on and the high and low priority alarm indicators (yellow and red lights) will not turn on or off or flickers.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON CONCHA NEPTUNE
Type of Device
RESPIRATORY GAS HUMIDIFIER
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key5939677
MDR Text Key54329954
Report Number3003898360-2016-00857
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number425-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/30/2016
Initial Date FDA Received09/09/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/06/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/14/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-