Model Number XPO100 |
Device Problems
No Audible Alarm (1019); Out-Of-Box Failure (2311)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional information become available, a supplemental record will be filed.
|
|
Event Description
|
Dealer is stating out of box failure, that there is no breath alarm will not activate, unit goes into an auto pulse mode without activation.
|
|
Manufacturer Narrative
|
Additional/updated information was added to reflect the device being returned to the manufacturer for evaluation.The result of the evaluation was that the restrictor tube was blocked by valves and that was causing the breath alarm to not function, which confirmed the original complaint issue.
|
|
Event Description
|
Dealer is stating out of box failure, that there is no breath alarm will not activate, unit goes into an auto pulse mode without activation.
|
|