• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP, SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP, SYSTEM 1 Back to Search Results
Model Number 801041
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/22/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Evaluation is in progress, but not yet concluded.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the roller pump display went in and out (blank).The user was still able to control using central control monitor (ccm) but changed out unit before case.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was not confirmed.During laboratory analysis, the reported complaint could not be duplicated.The product surveillance technician (pst) observed no blanking out of the display throughout testion conditions.Customer product was sent to service to be brought to manufacturers specifications before being returned to clinical use.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO ADVANCED PERFUSION SYSTEM 1
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP, SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key5948940
MDR Text Key54679799
Report Number1828100-2016-00615
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131618
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801041
Device Catalogue Number801041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/22/2016
Initial Date FDA Received09/14/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/14/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/14/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-