Model Number P0800010019
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem Pressure Sores (2326)
Event Date 08/19/2016
Event Type  Injury  
Manufacturer Narrative
The hill-rom technician found the microspheres were not moving properly.The technician moved the microspheres around and massaged them to resolve the issue.Based on this information, no further action is required.The patients mother reported that the patient developed a new wound on the left ischium tuberosity that was blackened and eschar.The presence or eschar indicates a stage 3 or greater wound.Development of pressure ulcers is multifactorial and cannot be only attributed to performance of the surface.Risk factors include protein-calorie malnutrition, microclimate (skin wetness caused by sweating or incontinence), diseases that reduce blood flow to the skin, such as arteriosclerosis, or diseases that reduce the sensation in the skin, such as paralysis or neuropathy.Position changes are key to pressure sore prevention and treatment.These changes need to be frequent, repositioning needs to avoid stress on the skin, and body positions need to minimize the risk of pressure on vulnerable areas.
 
Event Description
Hill-rom received a report from the mother of the patient alleging new skin breakdown.The bed was located at the account.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITRON RITEHITE C-8 ASSY
Type of Device
BED, AIR FLUIDIZED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key5950414
MDR Text Key54700658
Report Number1824206-2016-00392
Device Sequence Number1
Product Code IOQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K964223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Patient Family Member or Friend
Type of Report Initial
Report Date 08/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other Caregivers
Device Model NumberP0800010019
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/19/2016
Initial Date FDA Received09/14/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age68 YR
Patient Weight73
-
-