• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRAEGER, INC. DRAEGER JAUNDICE METER JM-105; JM-105 JAUNDICE METER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRAEGER, INC. DRAEGER JAUNDICE METER JM-105; JM-105 JAUNDICE METER Back to Search Results
Model Number JM-105
Device Problem Display Difficult to Read (1181)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/12/2016
Event Type  Injury  
Event Description
"critical" transcutaneous billrubin level on this device (greater than 340 umol/l or greater than 20 mg/dl) registers as flashing "- 0 -" with units displayed below (either mg/dl or umol/l).On more than one occasion nurses have taken this to literally mean "0" and have not acted appropriately (which is to either draw a serum bilirubin or at least notify the physician).This has occurred in spite of fairly intensive nursing training in device use.To avoid misinterpretation it is suggested that software be modified so that critical values are displayed as "hi", "- ! -" or similar.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRAEGER JAUNDICE METER JM-105
Type of Device
JM-105 JAUNDICE METER
Manufacturer (Section D)
DRAEGER, INC.
3135 quarry rd.
telford PA 18969
MDR Report Key5952373
MDR Text Key54935724
Report NumberMW5064767
Device Sequence Number1
Product Code MQM
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Physician
Type of Report Initial
Report Date 09/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberJM-105
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/13/2016
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-