Dexcom was made aware on (b)(4) 2016, that on (b)(6) 2016 the receiver display screen went white.There was no report of any injury or medical intervention.No additional event or patient information is available.Complaint device was returned for evaluation.A visual exterior inspection was performed on the receiver and it passed.A global receiver functional test was performed on the receiver and it passed the lcd test.Receiver's data logs were downloaded and reviewed, finding a screen error alarm, in addition to a firmware and a hardware error.Based on the finding of a firmware error this is being reported.The complaint was confirmed.The root cause could not be determined post investigation.
|