Model Number 250
Device Problems Disconnection (1171); Device Stops Intermittently (1599); Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/25/2016
Event Type  malfunction  
Event Description
It was reported by the company representative that the physician's serial cable was causing intermittent communication issues with the programming tablet.It was confirmed that troubleshooting, such as unplugging the serial cable and plugging it back in and testing on a known good programming tablet, were performed.It was found that only replacing the cable would resolve the issue.The serial cable does not require return, as the failure mode is understood to be a failure of the serial cable associated with a disconnected wire connection.
 
Manufacturer Narrative
Device manufacture date; corrected data: this information was inadvertently reported incorrectly on the initial mfr.Report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5953918
MDR Text Key55211130
Report Number1644487-2016-02089
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number250
Other Device ID NumberVERSION 11.0.5
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Other
Initial Date Manufacturer Received 08/25/2016
Initial Date FDA Received09/15/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/03/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/21/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N