• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES SPINE 5.5 EXP VERSE DI SET SCREW; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES SPINE 5.5 EXP VERSE DI SET SCREW; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Catalog Number 199721000
Device Problem Torn Material (3024)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/22/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A complaint investigation will be performed.The complaint product is not available for the investigation.A supplemental report is not anticipated unless the results of the complaint investigation identify a corrective action or additional relevant information.Should the product become available, a physical evaluation will be conducted and a supplemental report filed with the results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Scoliosis surgery: while correcting the hyper kyphosis cross threading occurred and the threads of the screw head were damaged.We had to break off the tabs to place the dual innie using the approximator.Dr.Schenk did not want to remove the complete construct and tried so long until he could lock the innie in the damaged screw head.The first threads of the screw broke and the threads below were intact and the innie could be locked.During that maneuvers 8 innies have been damaged - metal spreads came out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 EXP VERSE DI SET SCREW
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
DEPUY SYNTHES SPINE
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle 02767
SZ   02767
Manufacturer Contact
jason busch
325 paramount drive
raynham, MA 02767
5088808201
MDR Report Key5954150
MDR Text Key54885934
Report Number1526439-2016-10672
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeAU
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 08/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number199721000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Hospital
Initial Date Manufacturer Received 08/22/2016
Initial Date FDA Received09/15/2016
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age16 YR
-
-