IRIS INTERNATIONAL ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEM; AUTOMATED URINALYSIS SYSTEM Back to Search Results
Model Number N/A
Device Problems Leak/Splash (1354); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/25/2016
Event Type  malfunction  
Manufacturer Narrative
The field service engineer (fse) was at the customer site and inspected the connectors for the color gravity module (cgm) tubing.The leak was caused by a tear in the cgm tubing (the tubing that connects from cgm to 2.5ml syringe).The fse replaced the tubing to resolve the issue.The repairs were verified per established service procedures.Bec internal identifier for this report is (b)(6).
 
Event Description
The customer reported that there was a contained fluid leak that consisted of a few milliliters (ml) of liquid in their ichem velocity automated urine chemistry system.The customer also reported experiencing control failures for multiple analytes.Erroneous patient results were not generated and there was no change or effect to patient treatment in connection to the event.The customer was wearing personal protective equipment (ppe), there was no direct exposure of mucous membranes.There was no injury or medical attention required.
 
Manufacturer Narrative
Supplemental was required to correct bec internal identifier.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEM
Type of Device
AUTOMATED URINALYSIS SYSTEM
Manufacturer (Section D)
IRIS INTERNATIONAL
9172 eton ave
chatsworth CA 91311
Manufacturer (Section G)
IRIS INTERNATIONAL
9172 eton ave
chatsworth CA 91311
Manufacturer Contact
gopal mohanty
9172 eton avenue
chatsworth, ca, CA 91311
8185277379
MDR Report Key5955202
MDR Text Key54889911
Report Number2023446-2016-00363
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101852
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number700-7177-001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/25/2016
Initial Date FDA Received09/15/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received09/28/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/22/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-