MAUDE Adverse Event Report: SYNTHES SELZACH 10MM CANNULATED TAPERED DRILL BIT; TEMPLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES SELZACH 10MM CANNULATED TAPERED DRILL BIT; TEMPLATE Back to Search Results
Catalog Number 03.037.021
Device Problem Break (1069)
Patient Problems Sedation (2368); Device Embedded In Tissue or Plaque (3165)
Event Date 09/01/2016
Event Type  Injury  
Manufacturer Narrative
Device was used for treatment, not diagnosis.Additional patient information: patient height reported as: (b)(6).Patient initials are (b)(6).Device is an instrument and is not implanted/explanted.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Manufacturing site: (b)(4), supplier: (b)(4), manufacturing date: 02.May 2016.No ncrs were generated during production.Review of the device history record showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that three drill bits broke and there was difficulty drilling through a trochanteric fixation nail advanced (tfna) during an intramedullary (im) rodding of a left hip.After the nail and guide wire had been placed, as the surgeon used a stepped drill, the back end of a stepped drill bit snapped off.Then another identical stepped drill bit broke at the tip.Another set was opened and a tapered drill bit broke at the tip when it hit the nail, leaving fragments in the patient's hip.When the surgeon could not drill through the tfna and into the femoral head, that original nail was removed and replaced with a trochanteric fixation nail (tfn).There was a reported one hour surgical delay and additional x-rays were required.The procedure was then successfully completed with no harm to the patient.Concomitant devices reported: stepped drill (part #unknown, lot #unknown, quantity 1).This is report 1 of 3 for (b)(4).
 
Manufacturer Narrative
A product development investigation was performed for the following received subject devices by customer quality (cq): one 10mm/130 deg ti cann tfna 320mm/left ¿ sterile (part number: 04.037.053, lot number: h127743), one 10mm cannulated tapered drill bit (part number: 03.037.021, lot number: f-19306), one 6mm/9mm cannulated stepped drill bit (part number: 03.037.022, lot number: f-19278), one 6mm/9mm cannulated stepped drill bit (part number: 03.037.022, lot number: f-19282).The complaint condition is confirmed for all parts.A visual inspection and drawing review were performed as part of this investigation.The returned parts were determined to be suitable for their intended use when employed and maintained as recommended.The complaint devices are part of the depuy synthes tfnadvanced proximal femoral nailing system for intramedullary fixation of proximal femoral fractures.The nail (part number: 04.037.053, lot number: h127743) was returned with significant damage along the outside wall of the proximal locking hole and the inner wall of the distal locking hole.Slight abrasion along one of the grooves of the nail was also identified; misalignment has been determined to be the reason why the damage occurred.The tapered drill bit (part number: 03.037.021, lot number: f-19306) was returned with a portion of the distal tip broken off (approximately 1.81 mm is missing from the device) and was not returned.The stepped drill bit (part number: 03.037.022, lot number: f-19278) was returned with the proximal tip broken off (approximately 20.37mm had sheared off) and was not returned.The stepped drill bit (part number: 03.037.022, lot number: f-19282) was returned with the distal tip broken off (approximately 4.68mm had sheared off) and was not returned.Relevant drawings for the returned devices were reviewed: the design, materials and finishing processes were found to be appropriate for the intended use of these devices.During the investigation no product design issues or discrepancies were observed that may have contributed to the complaint condition.A root cause could not be determined as the circumstances at the time of the event are unknown.During the investigation no product design issues or discrepancies were observed that may have contributed to the complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10MM CANNULATED TAPERED DRILL BIT
Type of Device
TEMPLATE
Manufacturer (Section D)
SYNTHES SELZACH
bohnackerweg 5
selzach CH254 5
SZ  CH2545
Manufacturer (Section G)
SYNTHES SELZACH
bohnackerweg 5
selzach CH254 5
SZ   CH2545
Manufacturer Contact
mark vornheder
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5959727
MDR Text Key55025693
Report Number3000270450-2016-10225
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.037.021
Device Lot NumberF-19306
Other Device ID Number(01)10886982070371(10)F-19306
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/01/2016
Initial Date FDA Received09/19/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/03/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/02/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
STEPPED DRILL, PART# UNKNOWN
Patient Outcome(s) Required Intervention;
Patient Age88 YR
Patient Weight72
-
-