DEPUY SYNTHES POWER TOOLS COMPACT SPEED REDUCER, 60:1; DRILLS, BURRS, TREPHINES & ACCESSORIES - COMPACT SPEED REDUCER Back to Search Results
Catalog Number CSR60
Device Problem Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).Initial reporter's phone number: (b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was duplicated and confirmed.The assignable root cause was determined to be due to wear from normal use and servicing over time.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported in the service order from (b)(6) that the compact speed reducer device had an oil leak.During service and evaluation, it was confirmed that the device had an oil leak.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPACT SPEED REDUCER, 60:1
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES - COMPACT SPEED REDUCER
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key5961107
MDR Text Key55423734
Report Number1045834-2016-12766
Device Sequence Number1
Product Code HBE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K042783
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 09/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCSR60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/09/2016
Initial Date FDA Received09/19/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/27/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-