MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS COLIBRI II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS COLIBRI II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.101
Device Problems Electrical /Electronic Property Problem (1198); Mechanical Problem (1384); Defective Device (2588); Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to device wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported by (b)(6) that during an unspecified surgical procedure, it was observed that the small battery drive device stopped rotating while drilling the bone; then emitted an unusual sound.During service and evaluation, it was determined that the motor on the device seized, was rough running and defective.It was not reported if there were any delays in the surgical procedure or if a spare device was available for use.There was no patient involvement reported.There were no reports of any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
Subsequent follow-up with the reporter, additional information was received.The reporter clarified that the statement "when i drill the bone" was reported in error and has been updated to "when i run the handpiece".Therefore, it was determined that the reported malfunction is unlikely to cause or contribute to serious injury or death if it were to recur.Therefore, this complaint is not reportable.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COLIBRI II
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
mark vornheder
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5963161
MDR Text Key55454449
Report Number8030965-2016-14911
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service and Testing Personnel
Type of Report Initial,Followup
Report Date 09/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/05/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/05/2016
Initial Date FDA Received09/20/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/03/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/14/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-