During a prophylactic generator replacement surgery, the surgeon saw the lead was tied in a knot.When he straightened the lead out, he noticed the insulation of the lead was degraded.Impedance on the previous device was less than 600ohms.Therefore the surgeon performed a full revision surgery.The explanted devices were received on 09/20/2016.Analysis is underway but has not been completed to date.
|
An analysis was performed on the returned lead portions and the reported "mechanical problem / abraded insulation" (lead section) allegation was verified.The electrodes were not returned for analysis; therefore, a complete evaluation could not be performed on the entire lead product.During the visual analysis the outer silicone tubing appeared to be compressed, twisted with abraded openings, in several areas.During the visual analysis, multiple lead portions appeared to be melted.What appeared to be spatter and pitting was found on the surface of the quadfilar coil strands.Based on the obvious signs of melting on the coil surfaces, it is possible the thermally-damaged coils were exposed to a high temperature device such as a cauterizing tool (electrosurgical unit) during the explant of this lead.The abraded openings and melted appearance found on the outer silicone tubing, most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing.For the observed inner tubing fluid remnants, there was no obvious path for fluid ingress other than the melted / torn ends that were made during the explanted process.The condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.No other obvious anomalies were noted.The setscrew marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, during the visual analysis, with no discontinuities identified.Various electrical loads were attached to the pulse generator and results of diagnostic tests demonstrate that accurate resistance measurements were obtained in all instances.A comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.The battery, 2.813 volts, shows an ifi=no condition.There were no performance or any other type of adverse conditions found with the pulse generator.
|