CYBERONICS, INC. LEAD MODEL 302 Back to Search Results
Model Number 302-20
Device Problems Degraded (1153); Knotted (1340); Low impedance (2285)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/25/2016
Event Type  malfunction  
Event Description
During a prophylactic generator replacement surgery, the surgeon saw the lead was tied in a knot.When he straightened the lead out, he noticed the insulation of the lead was degraded.Impedance on the previous device was less than 600ohms.Therefore the surgeon performed a full revision surgery.The explanted devices were received on 09/20/2016.Analysis is underway but has not been completed to date.
 
Event Description
An analysis was performed on the returned lead portions and the reported "mechanical problem / abraded insulation" (lead section) allegation was verified.The electrodes were not returned for analysis; therefore, a complete evaluation could not be performed on the entire lead product.During the visual analysis the outer silicone tubing appeared to be compressed, twisted with abraded openings, in several areas.During the visual analysis, multiple lead portions appeared to be melted.What appeared to be spatter and pitting was found on the surface of the quadfilar coil strands.Based on the obvious signs of melting on the coil surfaces, it is possible the thermally-damaged coils were exposed to a high temperature device such as a cauterizing tool (electrosurgical unit) during the explant of this lead.The abraded openings and melted appearance found on the outer silicone tubing, most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing.For the observed inner tubing fluid remnants, there was no obvious path for fluid ingress other than the melted / torn ends that were made during the explanted process.The condition of the returned lead portions is consistent with conditions that typically exist following an explant procedure.No other obvious anomalies were noted.The setscrew marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, during the visual analysis, with no discontinuities identified.Various electrical loads were attached to the pulse generator and results of diagnostic tests demonstrate that accurate resistance measurements were obtained in all instances.A comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.The battery, 2.813 volts, shows an ifi=no condition.There were no performance or any other type of adverse conditions found with the pulse generator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 302
Type of Device
LEAD
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5966266
MDR Text Key55626707
Report Number1644487-2016-02151
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 08/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/30/2009
Device Model Number302-20
Device Lot Number200364
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2016
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 08/25/2016
Initial Date FDA Received09/21/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/13/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age18 YR
-
-