Visual inspection confirmed the reported complaint.The device was received with a portion of the distal tip broken off.The broken tip was returned with the device.A visual evaluation of the device via 40x magnification identified impact marks on the broken end of the device.The impact marks are consistent with damage from a hard, sharp surface.A review of the device history records did not identify any additional discrepancies during manufacture.No additional complaints are on file for this lot and reported failure mode.There are no indications that would suggest that the device did not meet product specifications upon release into distribution.
|