Additional narrative: (b)(4).(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
It was reported by germany that during service and evaluation, it was discovered that the motor seized, was running rough and defective on the small battery drive device.It was further determined that the motor and control were defective.It was further determined during the pre-repair diagnostics assessment that the device failed for check the attachment coupling, check the battery case coupling, check the off/osc/on switch mode function, check the oscillation angle, check the triggers and ecu function, check switching function, check compatibility between colibri/sbd and colibri ii/sbd ii, check the function with epd-console and for check power at the motor with test bench.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|