DEPUY SYNTHES POWER TOOLS AIR REAMER/DRILL; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 511.606 |
Device Problems
Partial Blockage (1065); Break (1069); Defective Device (2588); Protective Measures Problem (3015)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.It was further determined that the gear was torn off.The assignable root cause was determined to be due to improper handling, which is user error, misuse, and / or abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported by (b)(6) that during service and evaluation, it was discovered that the motor seized, blocked, was running rough and defective on the air reamer/drill device.It was further determined that the housing was damaged and the gear was broken.It was further determined during the pre-repair diagnostics assessment that the device failed for check the attachment coupling, check for free movement, check the symmetry, check for excessive noise, check for air leak, check the power with test stand and for check the starting behavior.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|