APOLLO ENDOSURGERY, INC. LAP-BAND AP STANDARD WITH ACCESS PORT I; ADJUSTABLE GASTRIC BAND Back to Search Results
Model Number B-2240
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Abscess (1690); Unspecified Infection (1930); Pain (1994)
Event Date 07/12/2016
Event Type  Injury  
Manufacturer Narrative
Taper ii.Device evaluation summary: the device was returned to apollo for analysis, and visual examination confirmed the connector type as taper ii.Visual inspection noted yellow particles on the outer surface of the band tubing.Non-penetrating nicks were observed around the suture holes.A port leak test was performed and no leakage was observed.A fill tube test was performed and not blockage or resistance to flow was observed.Striated openings were observed on the band tubing, consistent with the surgical removal of the device.The events reported are physiological complications and analysis of the device generally does not assist apollo in determining a probable cause for these events.Device labeling addresses the event as follows: adverse events: it is important to discuss all possible complications and adverse events with your patient.Complications which may result from the use of this product include the risks associated with the medications and methods utilized in the surgical procedure, the risks associated with any surgical procedure and the patient's degree of intolerance to any foreign object implanted in the body.Infection can occur in the immediate post-operative period or years after insertion of the device.In the presence of infection or contamination, removal of the device is indicated.Other adverse events considered related to the lap-band system that occurred in fewer than 1% of subjects included: esophagitis, gastritis, hiatal hernia, pancreatitis, abdominal pain, hernia, incisional infection, infection, redundant skin, dehydration, gi perforation, diarrhea, abnormal stools, constipation, flatulence, dyspepsia, eructation, cardiospasm, hematemesis, asthenia, fever, chest pain, incision pain, contact dermatitis, abnormal healing, edema, paresthesia, dysmenorrhea, hypochromic anemia, band leak, cholecystitis, esophageal dysmotility, esophageal ulcer, esophagitis, port displacement, port site pain, spleen injury and wound infection.Warning: patients should be advised that the lap-band ap system is a long-term implant.Explant (removal)and replacement surgery may be indicated at any time.Medical management of adverse reactions may include explantation.Revision surgery for explantation and replacement may also be indicated to achieve patient satisfaction.
 
Event Description
Reported as: a patient with the lap-band system was reported to be "evaluated for an infected lap band port.Patient had a colon perforation a year ago, repaired laparoscopically as well as a recent lap chole." patient had some redness, pain and formed abscess.The patient had the port removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAP-BAND AP STANDARD WITH ACCESS PORT I
Type of Device
ADJUSTABLE GASTRIC BAND
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC.
1120 s capitol of texas hwy
bldg 1, ste 300
austin TX 78746
Manufacturer (Section G)
ALLERGAN
global park free zone
900 global park
la aurora de heredia, costa rica
CS  
Manufacturer Contact
laura leboeuf
1120 s capitol of texas hwy
bldg 1, ste 300
austin, TX 78746
5122795141
MDR Report Key5971894
MDR Text Key55421047
Report Number3006722112-2016-00298
Device Sequence Number1
Product Code LTI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 08/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/02/2011
Device Model NumberB-2240
Device Catalogue NumberB-2240
Device Lot Number1779878
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/26/2016
Initial Date FDA Received09/22/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/04/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age52 YR
Patient Weight93
-
-