SMITH & NEPHEW, INC. TRUEPASS SUTURE PASSER SELF-CAPTURE; ACCESSORIES, ARTHROSCOPIC Back to Search Results
Catalog Number 72203791
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/22/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Visual inspection and functional testing could not be performed because the device in question was not returned for evaluation.Thus, the complaint could not be verified, nor could the root cause be determined with confidence.There are no indications that would suggest the device did not meet product specifications upon release into distribution.
 
Event Description
It was reported the truepass suture passer self-capture spring of the distal tip came out during the procedure.No patient injury or other complications were reported.
 
Manufacturer Narrative
Visual inspection confirmed that the torsion spring has been dislodged from the top jaw¿s spring insertion groove.After the evaluation the root cause for the reported issue was determined to be a design fault with the device.A review of the device history record was performed which confirmed no inconsistencies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEPASS SUTURE PASSER SELF-CAPTURE
Type of Device
ACCESSORIES, ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer Contact
jim gonzales
150 minuteman road
andover, MA 01810
5123585706
MDR Report Key5974883
MDR Text Key55895503
Report Number1219602-2016-00976
Device Sequence Number1
Product Code NBH
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 04/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72203791
Device Lot Number27870
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/25/2016
Initial Date FDA Received09/23/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/14/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-