A follow-up call with the doctor was done.The doctor informed millar of the following.Dr.(b)(6) called to inform a chest x-ray was performed on the patient and 5mm of the remaining catheter was located by the patient's fascia near the lvad.The tip appears to be 4cm deep in the skin.The patient has requested not to intervene to remove the portion of remaining catheter.Millar is unable to perform an evaluation on the catheter because it will not be returned.The catheter was discarded after the procedure, according to the doctor.Millar did perform a review of the device history records, which showed that each manufacturing and inspection operation was performed and indicated complete in accordance with the mikro-cath specifications and procedures.Based on the information received, the cause of the reported incident could not be conclusively determined.Device not returned.Device discarded.
|
The mikro-cath was placed in the left atrium of the patient during o.R placement of a left ventricular assist device lvad.The mikro-cath was passed through the chest wall using a peel-away long needle catheter and placed in the left atrium via access through the right superior pulmonary vein.Using sterile technique, the catheter was pulled from the chest.During the removal of the catheter, the tip broke off somewhere in the skin/soft tissue, as reported by dr.(b)(6).
|