Visual assessment of the device confirmed the reported breakage.The upper jaw has broken off.The tabs that interface with the shaft have broken off of the upper jaw causing the jaw to come free from the shaft.The shaft is bent downward.The condition of the device is consistent with excessive force being applied during use.Per the devices ifu under ¿precautions¿ ¿as with any surgical instrument, careful attention should be exercised to ensure that excessive force is not placed on the instrument.Excessive force can result in instrument failure.¿ no root cause related to the manufacture of the device can be established.No further investigation is required at this time.
|