• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERGE HEALTHCARE MERGE HEMODYNAMICS; PROGRAMMABLE DIAGNOSTIC COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERGE HEALTHCARE MERGE HEMODYNAMICS; PROGRAMMABLE DIAGNOSTIC COMPUTER Back to Search Results
Model Number MERGE HEMODYNAMICS 6.10
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Code Available (3191)
Event Date 08/25/2016
Event Type  malfunction  
Manufacturer Narrative
Merge technical support assisted the customer with preliminary troubleshooting and found that the server would not boot-up into "bios" or diagnostics.The site was then instructed to ship the faulty server to merge healthcare for repair.Upon evaluating the faulty server, a bad memory stick of ram was found and subsequently replaced along with the addition of 1.5 gb memory.The server passed all testing and was returned to the customer.Once re-installed, the server would boot-up and the proper light indicators were lit on the hard drive and led's to the mother board and power supplies; however, there was no video or mouse/keyboard control.Technical support instructed the user to re-seat the cpu (central processing unit).The customer then confirmed that the server was functioning as intended.(b)(4).
 
Event Description
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On (b)(6) 2016, a customer reported to merge healthcare that a memory error occurred on the site's server that caused the lab to be "down for a week" and subsequently caused procedures to be rescheduled during that time.With the merge hemo server not functioning as intended, there is a potential for delay in treatment that could result in harm to the patient.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE HEMODYNAMICS
Type of Device
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
michael diedrick
900 walnut ridge drive
hartland, WI 53029
2629123570
MDR Report Key5976144
MDR Text Key55568540
Report Number2183926-2016-00737
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082421
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 08/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMERGE HEMODYNAMICS 6.10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/25/2016
Initial Date FDA Received09/25/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-