Merge technical support assisted the customer with preliminary troubleshooting and found that the server would not boot-up into "bios" or diagnostics.The site was then instructed to ship the faulty server to merge healthcare for repair.Upon evaluating the faulty server, a bad memory stick of ram was found and subsequently replaced along with the addition of 1.5 gb memory.The server passed all testing and was returned to the customer.Once re-installed, the server would boot-up and the proper light indicators were lit on the hard drive and led's to the mother board and power supplies; however, there was no video or mouse/keyboard control.Technical support instructed the user to re-seat the cpu (central processing unit).The customer then confirmed that the server was functioning as intended.(b)(4).
|
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On (b)(6) 2016, a customer reported to merge healthcare that a memory error occurred on the site's server that caused the lab to be "down for a week" and subsequently caused procedures to be rescheduled during that time.With the merge hemo server not functioning as intended, there is a potential for delay in treatment that could result in harm to the patient.(b)(4).
|