Model Number 722026
Device Problem Unintended System Motion (1430)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
When the investigation as been completed, philips will inform the fda.(b)(4).
 
Event Description
Philips received a complaint from the customer in which was stated that they were unable to start the room for an emergency procedure.After a restart the c-arc made some unintended movements and collided into the wall of the room.The customer did not report any patient or user injury due to this problem.
 
Manufacturer Narrative
Investigation of the logfile showed that while the system was booting up, the customer powered down the system when it was almost started up completely.It also shows that certain functions are activated multiple times, when this is not required and seems almost erratic.The enmv (enable movement) error indicates a command was executed while the system was still starting up.The system was checked but there was no malfunctioning of the system that caused this problem.The system was monitored for a certain period and the issue had not occurred again.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD10
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS HEALTHCARE
veenpluis 4-6
p.o. box 10.000
best 5680 DA
NL  5680 DA
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
dusty leppert
veenpluis 4-6
p.o. box 10.000
best 5680 -DA
NL   5680 DA
MDR Report Key5976817
MDR Text Key55916654
Report Number3003768277-2016-00089
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K031333
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 09/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number722026
Device Catalogue Number722026
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/05/2016
Initial Date FDA Received09/26/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/03/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-