It was reported on (b)(6) 2016 that the physician's tablet was having trouble communicating with a 0 bytes message relaying on the screen.The tc performed troubleshooting and verified that the wand battery stayed on over 25 seconds.He verified that the tablet was not connected to the wall.He did not try removing and re-inserting the serial cable.The physician tried re-positioning the wand over the generator and the wand was away from other sources of electrical equipment.The patient being interrogated left without a successful programming but will come back for follow-up.The physician was adamant that the patient's device was not faulty or at end of service because he had just interrogated her successfully shortly prior to that visit and everything was working fine.The physician was given an extra black serial cable to see if that may resolve the issue at the next appointment for a patient.The black cable did resolve the issues.The serial cable does not require return, as the failure mode is understood to be a failure of the serial cable associated with a disconnected wire connection.
|