Model Number MODEL 250
Device Problems Communication or Transmission Problem (2896); Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/01/2016
Event Type  malfunction  
Manufacturer Narrative
 
Event Description
It was reported on (b)(6) 2016 that the physician's tablet was having trouble communicating with a 0 bytes message relaying on the screen.The tc performed troubleshooting and verified that the wand battery stayed on over 25 seconds.He verified that the tablet was not connected to the wall.He did not try removing and re-inserting the serial cable.The physician tried re-positioning the wand over the generator and the wand was away from other sources of electrical equipment.The patient being interrogated left without a successful programming but will come back for follow-up.The physician was adamant that the patient's device was not faulty or at end of service because he had just interrogated her successfully shortly prior to that visit and everything was working fine.The physician was given an extra black serial cable to see if that may resolve the issue at the next appointment for a patient.The black cable did resolve the issues.The serial cable does not require return, as the failure mode is understood to be a failure of the serial cable associated with a disconnected wire connection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5978175
MDR Text Key55941627
Report Number1644487-2016-02200
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 09/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 11.0.5
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 09/01/2016
Initial Date FDA Received09/26/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-