Catalog Number K08-MP5159A
Device Problems Device Packaging Compromised (2916); Packaging Problem (3007)
Patient Problem No Patient Involvement (2645)
Event Date 09/09/2016
Event Type  malfunction  
Manufacturer Narrative
The distributor returned 119 units for evaluation.A sample of units were examined visually.The complaint is confirmed.The root cause is attributed to the manufacturing process.The device history record was reviewed and no exception documents were found.The complaint database was reviewed and no similar complaints for this lot number were found.Corrective actions are in process for this failure mode.Merit determined on (b)(6) 2016 that a product removal would be required for four lots of kits affected by this packaging defect.Communications to consignees began on (b)(6) 2016.A report to the fda in accordance with 21 cfr 806.10 is in process and will be sent within the ten day requirement of initiating field action.No additional 3500a forms will be filed for this event.
 
Event Description
The distributor reported a defect in the packaging.This was identified during their initial inspection of received product.The devices were not sent to a user facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERIT CUSTOM KIT
Type of Device
CUSTOM KIT
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS, INC
1600 west merit parkway
south jordan UT 84095
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS, INC
1600 west merit parkway
south jordan UT 84095
Manufacturer Contact
casey hughes, ms, cqe
1600 west merit parkway
south jordan, UT 84095
8013164932
MDR Report Key5980490
MDR Text Key55739452
Report Number1721504-2016-00189
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K913682
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date 09/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2019
Device Catalogue NumberK08-MP5159A
Device Lot NumberH996433
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2016
Initial Date Manufacturer Received 09/09/2016
Initial Date FDA Received09/27/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/12/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction Number1721504-09/22/16-005R
Patient Sequence Number1
-
-