Catalog Number XMAX
Device Problems Defective Device (2588); Naturally Worn (2988); Protective Measures Problem (3015)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/26/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to improper handling.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported by (b)(6) that during service and evaluation, it was observed that the motor device had damaged locking components, a worn out locking mechanism and damaged housing.It was further observed that the device failed pre-test for cutter lock, safety temperature, sound and rotations per minute (rpm).This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMAX 2 PLUS MOTOR
Type of Device
MOTOR, DRILL, ELECTRIC - HANDPIECE
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key5984180
MDR Text Key56347768
Report Number1045834-2016-12849
Device Sequence Number1
Product Code HBC
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K080802
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 08/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberXMAX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/16/2016
Initial Date FDA Received09/28/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/25/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
MOTOR DEVICE
-
-